ACQUITY UPLC H-Class runs at a lower pressure than normal - WKB56050

Article number: 56050

SYMPTOMS

  • Typical run pressure is 2,000–4,000 psi lower than normal
  • System fails dynamic leak test only when the needle and seat are included
  • Needle seal readiness test fails

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY PREMIER
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

The needle seat has worn and no longer makes a good seal.

FIX or WORKAROUND

  1. Replace the needle and needle seat.
  2. Run z axis calibration, needle seal characterization, and the needle seal readiness check after replacement.

ADDITIONAL INFORMATION

Sample needle part numbers

  • 15uL H-Class: 700005215
  • 15uL H-Class Bio: 700005421
  • 15uL H-Class Plus (Textured): 700011628
  • 15uL H-Class Plus Bio (Textured): 700011658
  • 15uL ACQUITY Premier HPS (Textured): 700012820
  • 15uL ACQUITY Premier HPS: 700012821

Vespel seat Part number: 405011492

Seat port assembley part numbers:

  • Seat Port Assembly, .004 ID: 700005234
  • Seat Port Assembly, .004 ID, Bio: 700005682
  • Seat Port Assembly, HPS: 700012823