Carryover seen with an ACQUITY I-Class SM-FL - WKB56017
Article number: 56017
SYMPTOMS
- Carryover observed when running an analysis on SM-FL
ENVIRONMENT
- ACQUITY UPLC I-Class SM-FL
- ACQUITY UPLC
CAUSE
The inlet method uses the default injection mode Partial Loop Needle Overfill (PLNO).
FIX or WORKAROUND
Switch the injection mode to Partial Loop Pressure Assist (PLPA).
ADDITIONAL INFORMATION
- Switching from PLNO to PLPA injection mode can give certain applications significant carryover improvements.
- For more information, see the ACQUITY UPLC Sample Manager Injection Mode Quick Reference Card (715002108)
- Switching injection modes can cause significant changes to chromatography. Refer to the reference card before changing injection modes.