What are the UV cell options and wetted materials? - WKB80653
ENVIRONMENT
- ACQUITY TUV
- ACQUITY PDA
ANSWER
205000612 - ACQUITY PDA Flow Cell, Stainless Steel, 5 mm, 1500 nL
This 5 mm path length flow cell is constructed of stainless steel and is suitable for HPLC and biopharmaceutical applications.
205000613 - ACQUITY PDA Flow Cell, Titanium, 5mm, 1500 nL
This 5 mm path length flow cell is constructed of titanium and is suitable for HPLC and biopharmaceutical applications.
205015017 - ACQUITY PDA Flow Cell, Analytical, 10 mm, 500 nL
The 10 mm analytical light-guided flow cell delivers sensitivity and resolution to ACQUITY UPLC applications. Low-dispersion performance is achieved with a 500 nL illuminated volume and 127-μm PEEK interconnect inlet/outlet tubing. This flow cell is for use with the ACQUITY UPLC PDA or the ACQUITY UPLC PDAeλ detectors and is included when either detector is ordered.
205015019 - ACQUITY PDA Flow Cell, High Sensitivity, 25 mm, 2400 nL
The high sensitivity flow cell enables high sensitivity detection due to an extended 25-mm path length (2.4 μL illuminated volume). Low-dispersion performance is achieved with a 25-mm path length, 2400-μL volume, and 62.5-μm PEEK interconnect inlet/outlet tubing. This flow cell is for use with either the ACQUITY UPLC PDA or the ACQUITY UPLC PDAeλ detector.
TUV Flow Cell Options
205000610 - ACQUITY TUV Flow Cell, Stainless Steel, 5 mm, 1500 nL
This 5 mm path length flow cell is constructed of stainless steel and is suitable for HPLC and biopharmaceutical applications.
205000611 - ACQUITY TUV Flow Cell, Titanium, 5mm, 1500 nL
This 5 mm path length flow cell is constructed of titanium and is suitable for HPLC and biopharmaceutical applications.
205015016 ACQUITY TUV Flow Cell, Analytical, 10 mm, 500 nL
The 10 mm analytical light-guided flow cell delivers sensitivity and resolution to ACQUITY UPLC applications. Low-dispersion performance is achieved with a 500-nL illuminated volume and 127-μm PEEK interconnect inlet/outlet tubing. This flow cell is for use with the ACQUITY UPLC TUV detector and is included when the detector is ordered.
205015018 - ACQUITY TUV Flow Cell, High Sensitivity, 25 mm, 2400 nL
The high sensitivity flow cell enables high sensitivity detection due to an extended 25-mm path length (2.4-μL illuminated volume). Low-dispersion performance is achieved with a 25-mm path length, 2400-μL volume, and 62.5-μm PEEK interconnect inlet/outlet tubing.
ADDITIONAL INFORMATION
The only 10-mm flow cell is the Light Guided flow cell.
The only Light Guided flow cell is the 10-mm cell.