"Active pre-heater configured but not detected" message in the ACQUITY console - WKB80613

Article number: 80613

SYMPTOMS

  • "Active pre-heater configured but not detected" message in the ACQUITY console
  • Column was just changed

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS

CAUSE

The column preheater was not securely connected to the column heater.

FIX or WORKAROUND

Ensure that the preheater is securely connected to the column heater.

ADDITIONAL INFORMATION

 