Error message "seals not adjusted" during compression check with 2695 module - WKB80595
Article number: 80595
SYMPTOMS
- During compression check, the error message "seals not adjusted" appears, followed by "Compression check failed"
ENVIRONMENT
- 2695 Separations Module
- Alliance HPLC System
CAUSE
System not purged correctly
FIX or WORKAROUND
- Reboot the instrument.
- Run several (30) injector purges without compression check.
- Use a 100% organic mobile phase.
- Run another injector purge with compression check.
ADDITIONAL INFORMATION