Article number: 20632

ENVIRONMENT

  • Detectors with light-guided flow cells (TUV, PDA, FLR)

ANSWER

The best practice is to flush the system with 50:50 methanol/water, and then disconnect the flow cell outlet and install a pin plug. Do the same for the flow cell inlet. Disconnect all wires and the power source once the detector is powered-off. 

ADDITIONAL INFORMATION