How do you prepare an ACQUITY detector for long-term storage? - WKB20632
Article number: 20632
ENVIRONMENT
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC PDA eλ Detector
- ACQUITY UPLC FLR Detector
ANSWER
Extinguish the lamp
Flush the flow cell with water
Flush thew flow cell with 100% organic solvent such as methanol or acetonitrile
Power off the system
Cap the inlet and outlet ports of the flow cell
ADDITIONAL INFORMATION