High pressure on Alliance (rebuild cell) - WKB205640
Article number: 205640
SYMPTOMS
- High pressure alarm with column on the system
ENVIRONMENT
- Alliance
CAUSE
Partially blocked cell
FIX or WORKAROUND
Inspect the pressure with the cell of the detector and compare with the pressure on another system in the same condition. If the difference between the two systems is more than double and flushing the cell does not resolve the problem, you must rebuild the cell (WAS081346).
ADDITIONAL INFORMATION
- High pressure on Alliance (replace filter) - WKB87851
- High pressure on Alliance (blocked injector fluidics)
- High pressure on Alliance (flush cell)