High pressure on Alliance (rebuild cell) - WKB205640

Article number: 205640

SYMPTOMS

  • High pressure alarm with column on the system

ENVIRONMENT

  • Alliance

CAUSE

Partially blocked cell

FIX or WORKAROUND

Inspect the pressure with the cell of the detector and compare with the pressure on another system in the same condition. If the difference between the two systems is more than double and flushing the cell does not resolve the problem, you must rebuild the cell (WAS081346).