High pressure on Alliance (blocked injector fluidics) - WKB205636
Article number: 205636
SYMPTOMS
- High pressure alarm with column on the system
ENVIRONMENT
- Alliance
CAUSE
Partially blocked injector fluidics
FIX or WORKAROUND
Disconnect the inlet of the column. Compare the pressure with another system under the same conditions (plumbing and composition). If there is a big difference, a blockage is present in the fluidics of the injector.
ADDITIONAL INFORMATION
- High pressure on Alliance (replace filter) - WKB87851
- High pressure on Alliance (flush cell)
- High pressure on Alliance (rebuild cell)