Peaks not found error, lamp lighting error with ACQUITY TUV - WKB205540
Article number: 205540
SYMPTOMS
- Peaks not found error on startup of TUV detector
- Detector lamp lighting failure/error
- New lamp
- Low energy counts at 230 nm
ENVIRONMENT
- ACQUITY UPLC Tunable UV Detector
- ACQUITY Arc
- ACQUITY Arc Premier
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC PDA Detector
- ACQUITY Premier System
CAUSE
Dirty or damaged flow cell
FIX or WORKAROUND
Clean the flow cell and power cycle the detector. If unsuccessful, replace the flow cell.
ADDITIONAL INFORMATION
Install a shunt flow cell (700002825), try initializing the TUV, and check the energy counts. If it initializes without errors and energy increases, replace the flow cell (205015016).