"Needle arm home h/w fault" on ACQUITY Sample Manager-FTN - WKB205698

Article number: 205698

SYMPTOMS

  • "Needle arm home h/w fault" on ACQUITY Sample Manager-FTN

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC Sample Manager - Flow Through Needle 

CAUSE

A misshaped needle caught on the rear of the needle theta drive, preventing movement.

FIX or WORKAROUND

Replace the sample needle.

ADDITIONAL INFORMATION

 