"Needle arm home h/w fault" on ACQUITY Sample Manager-FTN - WKB205698
Article number: 205698
SYMPTOMS
- "Needle arm home h/w fault" on ACQUITY Sample Manager-FTN
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC Sample Manager - Flow Through Needle
CAUSE
A misshaped needle caught on the rear of the needle theta drive, preventing movement.
FIX or WORKAROUND
Replace the sample needle.
ADDITIONAL INFORMATION
id205698, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+