Temperature Control Module II does not reach the programmed temperature - WKB20050

Article number: 20050

SYMPTOMS

  • A temperature is programmed for the TCM
  • The TCM communicates and the the programmed temperature is displayed 
  • The actual temperature of the Column Heater Module does not change

ENVIRONMENT

  • Empower 3
  • Temperature Control Module II with Column Heater Module
  • ACQUITY UPLC system

CAUSE

The Column Heater Module is not plugged in.

FIX or WORKAROUND

When Waters ships a new TCM/CHM, there is a warning label placed over the outlet where the power cable should plug in (see attachment). Ensure that this sticker is removed and the unit is plugged in during installation.