2695 Column temperature unable to reach set point - WKB200566
Article number: 200566
SYMPTOMS
- Column temperature is set to 35℃, but it remains as 28℃
- unable to reach set point
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
The column heater door is not tightly closed.
FIX or WORKAROUND
- Close the door until you hear a click sound.
ADDITIONAL INFORMATION