QDa leak in the middle of the probe tube - WKB200491

Article number: 200491

SYMPTOMS

  • Leak in the middle of the probe tube

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

The probe tube was caught in the front cover.
clipboard_eefe8157d51566febb86145378331c444.png

FIX or WORKAROUND

Replace the ESI probe assembly (700009641).

ADDITIONAL INFORMATION

Always confirm that the probe tube is inside (at the three red circles) before closing the front cover.
clipboard_e551555cc739942b266cf4b60c6c7c06a.png