Missed sample injections due to bent puncture needle with ACQUITY UPLC SM - WKB19278

Article number: 19278

SYMPTOMS

  • Instrument intermittently misses injections; misses are random, not method-specific
    • All re-injected samples were acceptable and consistent
  • New sample needle, sample loop, and injection valve
    •  30-µL steel needle
  •  All plate types and all positions show a random missed injection
    • Waters 96-1mL plates and Waters 2-mL vials

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • nanoACQUITY Sample Manager

CAUSE

Bent puncture needle led to variable vial and well plate punctures, and hence to missed injections.

FIX or WORKAROUND

  1. Replace puncture needle.