Ethernet instruments do not populate in the Waters DHCP server configuration window after power outage - WKB13958
Article number: 13958
SYMPTOMS
- PC reboot after power outage
- Clear Waters DHCP server configuration window
- Reboot instruments
- Instruments power-up properly but nothing appears in the Waters DHCP server configuration window
- Firewall is off
- Swapping Ethernet cables does not resolve
- Swapping Ethernet switch box does not resolve
ENVIRONMENT
- ACQUITY H-Class system
- MassLynx
- Empower 3
CAUSE
Instrument LAN card failure
FIX or WORKAROUND
- Replace and configure the instrument LAN card.
ADDITIONAL INFORMATION