With ACQUITY UPLC FLR Detector and 2475 Multi-Wavelength Fluorescence Detector, flat baseline observed during sample injections - WKB118075
Article number: 118075
SYMPTOMS
- Flat, no-noise baseline observed
- Low standards missing; high standards present
- Random baseline spike
ENVIRONMENT
- ACQUITY UPLC FLR Detector
- 2475 Multi-Wavelength Fluorescence Detector
CAUSE
The FLR detector has been powered-on, with the lamp off when not in use, for several weeks, since the last preventive maintenance (PM) or lamp change. Internal calibration and normalization diagnostics have not been performed since the PM.
FIX or WORKAROUND
- Remove the LC column and install a union/restrictor.
- Power cycle the FLR.
- Flush the flow cell with 100% LC/MS grade (or equivalent) water for 30 minutes at 1.0 mL/min.
- Perform wavelength calibration.
- Perform normalization diagnostic.
- Perform Raman signal-to-noise diagnostic.
ADDITIONAL INFORMATION
See also:
What is "normalized mode" on a 2475 or ACQUITY FLR? - WKB29076
FLR Detector - How often should you normalize? - WKB77456
id118075, 2475, 2475B, A-10FL, A-30FL, UPBSMIVD, UPFLRARC, UPFLRDET, UPIBSMIVD, UPLCCOLHTR, UPNBINARY, UPPFLR, UPSMIVD, UPSMPMGR