On ACQUITY H-Class, no liquid comes out of the waste line while flushing instrument - WKB114370

Article number: 114370

SYMPTOMS

  •  No liquid comes out of the waste line while flushing instrument

ENVIRONMENT

  •  ACQUITY H-Class

CAUSE

The waste cup is filled with debris.

The waste line is not cut to a reasonable length.

FIX or WORKAROUND

  1. Remove and clean the waste cup.
  2. Cut the waste line so that it has a straight drop to the waste bucket and is not coiled.

ADDITIONAL INFORMATION

 