Blue liquid comes out from the Reagent Manager - WKB11436

Article number: 11436

SYMPTOMS

  • High baseline observed on chromatogram
  • Blue liquid observed on waste line of Reagent Manager

ENVIRONMENT

  • Reagent Manager (RMA)

CAUSE

The curtain inside of the pulse dampener breaks and blue liquid leaks from the pulse dampener.

Capture.JPG

FIX or WORKAROUND

Replace the pulse dampener (700000147).

ADDITIONAL INFORMATION