No solvent flow from ACQUITY Solvent Manager - WKB10645
Article number: 10645
SYMPTOMS
- No solvent flow from solvent line(s).
- Very low and unstable pressure when running flow through X and/or Y solvent lines.
ENVIRONMENT
- ACQUITY UPLC Binary Solvent Manager (BSM)
- ACQUITY Arc Quaternary solvent Manager-R (QSM-R)
- ACQUITY QSM
- ACQUITY PREMIER BSM
CAUSE
No solvent passing through the pump heads due to dry solvent line(s).
FIX or WORKAROUND
- Thoroughly prime the solvent line(s).
- Ensure pressure is stable and as expected.
ADDITIONAL INFORMATION
Manual solvent line priming is often the case when pump had ran dry. See articles How do I manually prime an ACQUITY Binary Solvent Manager? - WKB6161 for BSM or article How do I manually prime the ACQUITY Quaternary Solvent Manager? - WKB6163 for a QSM.
id10645, A-30SDM, pressure fluctuation, UPAQSM, UPARQSM, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM, UPPQSM, UPQSM, UPQSM+, UPQSMBIO