Skip to main content
Waters

No solvent flow from ACQUITY Solvent Manager - WKB10645

Article number: 10645

SYMPTOMS

  • No solvent flow from solvent line(s). 
  • Very low and unstable pressure when running flow through X and/or Y solvent lines.

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY Arc Quaternary solvent Manager-R (QSM-R)
  • ACQUITY QSM
  • ACQUITY PREMIER BSM

CAUSE

No solvent passing through the pump heads due to dry solvent line(s).

FIX or WORKAROUND

  1. Thoroughly prime the solvent line(s).
  2. Ensure pressure is stable and as expected. 

ADDITIONAL INFORMATION

Manual solvent line priming is often the case when pump had ran dry. See articles How do I manually prime an ACQUITY Binary Solvent Manager? - WKB6161 for BSM or article How do I manually prime the ACQUITY Quaternary Solvent Manager? - WKB6163 for a QSM. 

Not able to find a solution? Click here to request help.