Error "The project does not have a corresponding folder on RAW data share" appears in the NuGenesis Archive Agent's log file - WKB77959
Article number: 77959
SYMPTOMS
- The Archive Agent (AA) module is running with a file capture template line that uses Empower Data Adapter
- The following error message appears in the AA activity log:
- The project <Empower project name> does not have a corresponding folder on RAW data share. The Folder <Empower data UNC path> does not exist.
Please add the project <Empower project name> to the SDMS Exceptions List."
- The project <Empower project name> does not have a corresponding folder on RAW data share. The Folder <Empower data UNC path> does not exist.
- The error is repeated for all Empower projects, regardless of which Empower project NuGenesis attempts from a specific Empower server
ENVIRONMENT
- NuGenesis 8 SR2
- NuGenesis Data Adapters Release 3
- Empower 2 or Empower 3 client
CAUSE
The Windows user configured to run the NuGenesis RPC Service on the File Capture server does not have access to the Empower RAW data share.
Without at least read access to the Empower RAW data share (Default: Waters_Project$), the AA cannot access the Empower data UNC path.
In this case the Project folder on the RAW data share does exist, but the NuGenesis FC server could not see it through the network.
FIX or WORKAROUND
Update permission on the Empower RAW data share (Waters_Projects$) to allow read access to the NuGenesis Service account user.