Results with special limits (<LOQ, <LOD, for example) are sent from the Historical Stability template to NuGenesis LMS without a valuation code - WKB75755

Article number: 75755

SYMPTOMS

  • When filling out the Historical Stability template for legacy stability studies, results with special limits, such as <LOQ or <LOD, are sent to LMS Sample Management without a valuation code
  • The results in LMS cannot be manually edited because the test is linked to a document
  • "Allow special limits" is active for the test
  • The request in LMS cannot be completed and released because one or more results have no valuation

ENVIRONMENT

  • NuGenesis 9 LMS
  • NuGenesis 8 LMS

CAUSE

A defect in the Historical Stability template

FIX or WORKAROUND

  1. Open the Historical Stability template in the template editor.
  2. Edit the results entry section and switch it to Design mode.
  3. Select the Query tab.
  4. Enter the word VALUATION in cell L4.
  5. Enter the following formula into cell L5:
    • =IF($K5 <> "", 1, "")
  6. Copy that formula into all cells, from L6 down to L404.
  7. Click the Formulas tab in the Ribbon, and then click Name Manager.
  8. Select the named range "myresult" and change its range to:
    • =Querry!$J$4:$L$404
  9. Close Excel, save and approve the template.
  10. Select the tests and create a new document. The new document will use the updated template and provide a valuation code for tests with results such as <LOQ and <LOD.

ADDITIONAL INFORMATION

This article assumes that the Historical Stability template has been expanded to 400 rows as per article WKB75107.

This update sets a valuation code of 1 for all results entered through this form. If it is necessary to have NOK valuations entered through this form, contact Waters Professional Services for assistance. Setting a valuation code of 0 (for NOK) instead of 1 in the hidden sheet requires additional logic and form configurations that cannot be covered within the scope of an article.