"Error 10048: Only one usage of each socket address . . . is normally permitted" occurs when restarting the NuGenesis VISION service - WKB64717
Article number: 64717
SYMPTOMS
- The following error occurs when restarting the NuGenesis VISION service:
- Windows could not start the NuGenesis VISION Server service on local computer
- Error 10048: Only one usage of each socket address (protocol/network address/port) is normally permitted
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
Another service is currently running and listening on TCP port 80.
FIX or WORKAROUND
Do not restart the NuGenesis VISION service on its own. Follow the directions in the linked article to cleanly restart the service.
ADDITIONAL INFORMATION
id64717, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG