Error "The socket connection was aborted" appears in the NuGenesis Archive Agent log file - WKB59041
SYMPTOMS
- No Empower data is archived into NuGenesis
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
CAUSE
The "WatersService" service was not running during the import of the project into Empower; or, the machine running the SDMS DataExchangeService cannot connect to the WatersService.
FIX or WORKAROUND
- Start "WatersService" on the Empower server
- Ensure that the firewall on the Empower server, if active, allows the network ports as prescribed in the Waters Informatics Infrastructure Reference Guide, as linked in article WKB5512
- If using a virtual machine, ensure that the machine running the SDMS DataExchange services is on the same network as the Empower server
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent log file "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":
Message........: 2019-02-19 14:30:42: Beginning execution of archive request 1.
Message........: 2019-02-19 14:30:42: Server: VDB, Project: Empower, Template: EmpClient.
Message........: 2019-02-19 14:30:43: Archive scan with no files archived for command id 1. {"VDB","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 2/19/2019 2:30:42 PM)(End Time = 2/19/2019 2:30:43 PM)(Total Time = 0h 0m 1s)}.
"E ","2019/02/19 02:31:28 PM ","5 ","5 ","The socket connection was aborted. This could be caused by an error processing your message or a receive timeout being exceeded by the remote host, or an underlying network resource issue. Local socket timeout was '7.22:59:59.9957375'. While extracting keys for file : \\VCL\SDMSarchive\EMPOWER3\Test 10800 Disabled\d10800.dat ","Waters.SDMS.EmpowerDAProxy ","ClsEmp2FAExtract ","ExtractKeys ","1 ",""
"E ","2019/02/19 02:31:27 PM ","5 ","5 ","Invalid procedure call or argument While extracting keys for file : \\servername\share name\EMPOWER3\Project Name\d10800.dat ","NG80EVSEmp2PROXY ","clsEmp2FAExtract ","ExtractKeys ","1 ",""
Message........: 2019-02-19 14:31:27: ExecuteTemplateCommand. Unable to extract keys using objiCommonKEE.iExtractKeys.
Message........: 2019-02-19 14:31:27: FillTransactionDO. (5): Invalid procedure call or argument.
Message........: 2019-02-19 14:31:27: EMC: FillTransactionDO failure.
Message........: 2019-02-19 14:31:27: Rollback: Performing rollback for D:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\3781E4AD42BA49622DAEA8ABBE75BFCD.
"M ","2019/02/19 02:31:27 PM ","0 ","0 ","Archived Project:[EMPOWER3.Test 10800 Disabled] "," "," "," ","0 ",""
Message........: 2019-02-19 14:31:27: ExecuteTemplateCommand: The group is rolled back because of errors.
Message........: 2019-02-19 14:31:27: Archive scan summary for command id 1: {"VDB","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"} {Summary : (Command ScanPath = \\VCL\SDMSarchive)(Total Scanned Files = 1)(Total Archived Files = 1)(Total Bytes = 5157)(Start Time = 2/19/2019 2:30:43 PM)(End Time = 2/19/2019 2:31:27 PM)(Total Time = 0h 0m 44s)(Errors = 1)}.
Message........: 2019-02-19 14:31:27: Next scheduled runtime for archive request 1 is 2019-02-19 14:40:41.