Error "Error encountered while invoking remote EVS Server objects. :The type initializer for '<Module>' threw an exception." appears in the NuGenesis Archive Agent log file - WKB59043
Article number: 59043
SYMPTOMS
- No Empower data is archived into NuGenesis SDMS
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters
- Empower 3
CAUSE
Windows updates require a restart on the remote machine (the "Remote DCOM server" machine name in the NuGenesis Archive template line).
FIX or WORKAROUND
Restart the machine.
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent's log file Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log:
"E ","DATE TIME","5 ","5 ","Error encountered while invoking remote EVS Server objects. :The type initializer for '<Module>' threw an exception.. ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: DATE TIME: EMC: PerformPreProcess error.