Error "PerformProjBackup failed" appears in the NuGenesis Archive Agent log file - WKB59046
SYMPTOMS
- No Empower data is archived into NuGenesis
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
CAUSE
The Empower user, as specified in the Archive Template Request Line, does not have the “Alter Any Project” or the “Backup Project” privileges.
FIX or WORKAROUND
In Empower, assign the appropriate privileges to the user.
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent's log file "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":
Message........: 2018-12-18 14:05:43: Beginning execution of archive request 1.
Message........: 2018-12-18 14:05:43: Server: NUG, Project: Empower, Template: EmpServer.
Message........: 2018-12-18 14:05:44: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:05:43 PM)(End Time = 12/18/2018 2:05:44 PM)(Total Time = 0h 0m 1s)}.
"E ","2018/12/18 02:05:47 PM ","-2147220896 ","-2147220896 ","-2147220896Exception from HRESULT: 0x80040260 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","BackupProjectTK ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","PerformProjBackup failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-SRV ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-12-18 14:05:47: EMC: PerformPreProcess error.
Message........: 2018-12-18 14:05:50: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:05:47 PM)(End Time = 12/18/2018 2:05:50 PM)(Total Time = 0h 0m 3s)}.
Message........: 2018-12-18 14:05:50: Next scheduled runtime for archive request 1 is 2018-12-18 14:06:32.