Error "Project Locking failed" appears in the NuGenesis Archive Agent log file - WKB59051
SYMPTOMS
- No Empower data is archived into NuGenesis
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
CAUSE
The Empower user, as specified in the Request Line of the File Capture Template, does not have the Empower “Lock Project” privilege.
FIX or WORKAROUND
In Empower, assign the appropriate privilege to the user.
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent log file "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":
Message........: 2018-12-18 14:31:44: Beginning execution of archive request 1.
Message........: 2018-12-18 14:31:44: Server: NUG, Project: Empower, Template: EmpServer.
Message........: 2018-12-18 14:31:45: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:31:44 PM)(End Time = 12/18/2018 2:31:45 PM)(Total Time = 0h 0m 1s)}.
"E ","2018/12/18 02:31:52 PM ","-2147220896 ","-2147220896 ","Exception from HRESULT: 0x80040260 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","LockProject ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","Project Locking failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-SRV ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-12-18 14:31:52: EMC: PerformPreProcess error.
"M ","2018/12/18 02:31:54 PM ","0 ","0 ","Archived Project:[EMP3.MANY\Tea Group\Tea 002] "," "," "," ","0 ",""
Message........: 2018-12-18 14:31:54: Archive scan summary for command id 1: {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"} {Summary : (Command ScanPath = \\WB-EMP-SRV\SDMSarchive)(Total Scanned Files = 14)(Total Archived Files = 14)(Total Bytes = 3570891)(Start Time = 12/18/2018 2:31:45 PM)(End Time = 12/18/2018 2:31:54 PM)(Total Time = 0h 0m 9s)(Errors = 0)}.
Message........: 2018-12-18 14:32:03: Next scheduled runtime for archive request 1 is 2018-12-18 14:32:32.