Error "Exception of type 'System.OutOfMemoryException' was thrown" appears in the NuGenesis Archive Agent log file - WKB59028
SYMPTOMS
- No Empower data is archived into NuGenesis 8 SR-2
ENVIRONMENT
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters Release 3 HF1
- Empower 3
CAUSE
Attempting to archive an Empower project with JCAMP conversion enabled. The targeted project is too large for the conversion module to handle.
FIX or WORKAROUND
Install DAR3-HF2 on the Empower Remote DCOM server to resolve this issue.
ADDITIONAL INFORMATION
The following messages appear in the Archive Agent's log file "Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log":
Message........: 2018-01-16 07:22:21: Beginning execution of archive request 1.
Message........: 2018-01-16 07:22:21: Server: NUG, Project: Empower, Template: EmpClient.
Message........: 2018-01-16 07:22:22: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 7:22:21 AM)(End Time = 1/16/2018 7:22:22 AM)(Total Time = 0h 0m 1s)}.
"E ","2018/01/16 08:01:04 AM ","-2147417851 ","-2147417851 ","The server threw an exception. (0x80010105 (RPC_E_SERVERFAULT)): ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","NGPerformConvTasks ","1 ",""
"E ","2018/01/16 08:01:18 AM ","7 ","7 ","Exception of type 'System.OutOfMemoryException' was thrown.:EMP3.CUSTOMERS\GSK\EVR_MV_SA_MDI_FP_FAST_08_03 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","ExtractKeys ","1 ",""
"E ","2018/01/16 08:01:18 AM ","0 ","0 ","Error extracting keys for the project:EMP3.CUSTOMERS\GSK\EVR_MV_SA_MDI_FP_FAST_08_03 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/01/16 08:01:18 AM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/01/16 08:01:18 AM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-CL ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-01-16 08:01:19: EMC: PerformPreProcess error.
Message........: 2018-01-16 08:01:19: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 7:22:22 AM)(End Time = 1/16/2018 8:01:19 AM)(Total Time = 0h 38m 57s)}.
Message........: 2018-01-16 08:01:19: Next scheduled runtime for archive request 1 is 2018-01-16 11:00:42.