Error "The project does not have a corresponding folder on RAW data share" appears in the NuGenesis Archive Agent log file - WKB49616
Article number: 49616
SYMPTOMS
- The Archive Agent (AA) module is running with a file capture template line that uses Empower Data Adapter
- The following error message appears in the AA activity log:
- The project [Empower project name] does not have a corresponding folder on RAW data share. The folder [Empower data UNC path] does not exist.
Please add the project [Empower project name] to the SDMS Exceptions List.
- The project [Empower project name] does not have a corresponding folder on RAW data share. The folder [Empower data UNC path] does not exist.
- The Empower project referenced in the error has a valid data folder in the Empower data UNC path with a ^# at the end
ENVIRONMENT
- NuGenesis 8 SR2
- NuGenesis Data Adpaters Release 3
- Empower 2 or Empower 3 client
CAUSE
The Empower project being captured had a name that duplicates that of an existing project.
Empower manages duplicate project names by creating a data folder with the project name and a ^1 or ^2 extension.
NuGenesis Data Adapters does not identify the Empower Data folder correctly.
FIX or WORKAROUND
Apply the NuGenesis Data Adapters 3 Hotfix 2.
ADDITIONAL INFORMATION
NuGenesis Data Adapters 3 Hotfix 1 contains the fix, but Hotfix 2 has the code changes that were made in Hotfix 1.