Error "Login to Millennium service failed" appears in the NuGenesis SDMS Archive Agent log - WKB26457

Article number: 26457

SYMPTOMS

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS

CAUSE

The Exceptions List username or password in the NuGenesis File Capture template line is not valid. 

FIX or WORKAROUND

Edit the NuGenesis File Capture template line to update the exceptions list to use a username and password combination that is valid in Empower.

ADDITIONAL INFORMATION

Example error message from the Archive Agent activity log:

Message : 4/25/2018 4:44:00 PM : Beginning execution of Archive request 2. 
"E ","2018/04/25 04:44:01 PM ","-1 ","-1 ","Login to Millennium service failed. ","ng70EVSEmp2Svr ","clsEmp2PreProc ","ValidateInputs ","1 ","" 
"E ","2018/04/25 04:44:01 PM ","-6 ","-6 ","Input Validation failed. ","ng70EVSEmp2Svr ","clsEmp2PreProc ","PerformPreProcess ","1 ","" 
"E ","2018/04/25 04:44:01 PM ","-1 ","-1 ","PreProcessor call failed for the remote server: servername ","ng70EVSEmp2PROXY ","clsMillPreProc ","PerformPreProcess ","1 ","" 
Message : 4/25/2018 4:44:01 PM : EMC : PerformPreProcess error. 
Message : 4/25/2018 4:44:01 PM : Archive scan with no files archived for command id 2.{"Waters Corp.","Empower-2 Projects","0"}. 
Message : 4/25/2018 4:44:01 PM : Next scheduled runtime for archive request 2 is 4/25/2018 7:30:00 PM.