Error "[NG14420] Operating System Error : Failed to open file" when restoring files from a NuGenesis OSM device - WKB26315
Article number: 26315
SYMPTOMS
- The following error occurs when restoring files from an OSM device:
- "[NG14420] Operating System Error : Failed to open file"
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
The OSM device configuration is incorrect. The device type is Hard Disk and the device's UNC path and drive letter do not match.
FIX or WORKAROUND
- Log in to the NuGenesis SCC Manager.
- Select the Offline Storage Manager module.
- Stop the module if it is currently started.
- Select the Device Configuration tab.
- Select the device name and click Edit.
- For a Hard Disk-type device, the Device Path (UNC) must be the UNC path for the Device Drive (Local) entry.
- Example: Device Drive = E:\
- Device Path = \\servername\$e OR an equivalent UNC path for the E: drive
- The NuGenesis RPC service account must have read, write, and delete permissions on the Device Drive and Device Path.
- For a RAID or HSM Store-type device, the Device UNC Path need not match the Device Drive; in fact, the OSM module will use the Device UNC path exclusively.
- Example: Device Drive = C:\
- Device Path = \\servername\OSMData