Error "The project does not have a corresponding folder on RAW data share" appears in the NuGenesis Archive Agent's log file for some Empower projects - WKB223086
Article number: 223086
SYMPTOMS
- The Archive Agent (AA) module is running with a file capture template line that uses Empower Data Adapter
- The following error message appears in the AA activity log:
- The project <Empower project name> does not have a corresponding folder on RAW data share. The Folder <Empower data UNC path> does not exist.
Please add the project <Empower project name> to the SDMS Exceptions List."
- The project <Empower project name> does not have a corresponding folder on RAW data share. The Folder <Empower data UNC path> does not exist.
- The error message repeats for the same project on each archive run
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
CAUSE
The project in Empower has an invalid/obsolete raw data share.
FIX or WORKAROUND
The project cannot be archived by NuGenesis SDMS. Edit the project exceptions list in the SDMS Archive Template to exclude this project from further archive runs.
ADDITIONAL INFORMATION