How to enable debug logging for the Empower data adapter in NuGenesis SDMS 7.1 - WKB10550
Article number: 10550
OBJECTIVE or GOAL
Activate debug logging for the "Remote" half of the NuGenesis SDMS 7.1 data adapter for Empower.
ENVIRONMENT
- NuGenesis SDMS 7.1
PROCEDURE
- On the SDMS server, stop the Archive Agent in the Service Configuration console.
- Restart the RPC Service.
- On the Empower system, rename "D:\Program Files\NuGenesis 7.1\EVS\ng70EVSEmp2Svr.exe" to "ng70EVSEmp2Svr.exe-orig" (path may be different).
- Copy the debug version (attached) of ng70EVSEmp2Svr.exe into D:\Program Files\NuGenesis 7.1\EVS (path may be different).
- On the SDMS server, start the Archive Agent in the Service Configuration Console.
- Find additional log files (EmpowerLog.log) in C:\Windows\system32.
ADDITIONAL INFORMATION
See also: the debug logging procedure for the Empower data adapter in SDMS 8.