Skip to main content
Waters

With which MassLynx 4.2 SCNs can Skyline be used to successfully export MRM methods? - WKB92770

Article number: 92770

ENVIRONMENT

  • MassLynx 4.2
  • Skyline
  • Xevo TQ-XS
  • Xevo TQ-S Micro
  • Xevo TQ-S

ANSWER

Empirical testing and real world usage show that when Skyline is installed alongside the following MassLynx 4.2 SCNs it is possible to export MassLynx methods from Skyline using an appropriate MS method template. More recent SCNs for the instruments listed below will also work, as will all SCNs for the Xevo TQ Absolute

 

MassLynx 4.2 Software Change Note Instrument

SCN982

Xevo TQ-XS

SCN977, SCN1001, SCN1013 Xevo TQ-S Micro
SCN997 Xevo TQ-S

ADDITIONAL INFORMATION

If you try to use Skyline to export MassLynx methods with the following MassLynx 4.2 SCNs, an error occurs, and no MS method is created.

MassLynx 4.2 Software Change Note Instrument
SCN 943 Xevo TQ-XS
SCN 985 Xevo TQ-D
SCN 986 Xevo TQ-S

 

id92770, MLYNX, MLYNXV41, SCN 1001, SCN 1013, SCN 977, SCN 982, SCN 997, SCN943, SCN985, SCN986, SUPMM, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.