Skyline method export error with Xevo TQ-XS and MassLynx 4.2 SCN 943 - WKB60478
Article number: 60478
SYMPTOMS
- Unable to export MassLynx methods from Skyline
- When selecting File > Export > Export method, the following error message appears:
- ERROR: Method not found: 'BooleanCompoundDatabaseClassLibrary.CompoundDatabase.InsertNewIonDetails(Int32, Int32, TransMode, Int32, Int32, Int32, Int32, Int32, Int32
- Problem occurs whether a fresh MS method template (.exp file) is created in MassLynx 4.2 SCN 943 or if one of the templates in C:/MassLynx/Default.pro/Acqdb is used
- Problem occurs when you export a single method or multiple methods
ENVIRONMENT
- Skyline
- Xevo TQ-XS
- MassLynx 4.2 SCN 943
CAUSE
Incompatibility between Skyline and MassLynx 4.2 SCN 943 (Xevo TQ-XS)
MassLynx 4.2 SCNs for other quad instruments are not affected.
FIX or WORKAROUND
Issue is resolved in MassLynx 4.2 SCN 982
- Update system to MassLynx 4.2 SCN 982.
- Create a new method template using MassLynx 4.2 SCN 982.
- Repeat method export from Skyline using the new template.
ADDITIONAL INFORMATION
Method export from Skyline may require a TargetLynx installation in MassLynx, and may not work if the "desktop processing" option was selected when MassLynx was installed.