MassLynx chromatogram TIC displays non-zero baseline before acquisition - WKB60433

Article number: 60433

SYMPTOMS

  • MassLynx chromatogram shows TIC response of 5e3 before sample acquisition starts

ENVIRONMENT

  • MassLynx

CAUSE

The Chromatogram display setting is incorrect, "Normalize Data To" is set to "Intensity" with a programmed baseline value of 5000.

FIX or WORKAROUND

  1. In Chromatogram view, go to Display > View.
  2. Select "Largest Peak" for "Normalize Data To:". 

ADDITIONAL INFORMATION