MS waiting for DHCP in Hyperterminal Log - WKB7608
Article number: 7608
SYMPTOMS
- MS does not receive communications after a restart
- Cannot ping EPC
- Nothing showing in WDHCPserver wizard - including ACQUITY modules
- Hyperterminal Boot Sequence hanging at "Waiting for DHCP" line:
boot device: gei
unit number: 0
processor number: 0
host name: bootHost
file name: vxWorks
flags (f): 0x40
target name (tn): vxtarget
startup script (s): script.txt
other (o): fei
Waiting for DHCP, CTRL-X to abort.
................................_
ENVIRONMENT
- MassLynx
- Xevo TQ-S (but any MS with Dynamic Boot sequence DHCP comms)
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Windows
- Antivirus software
CAUSE
Antivirus software is blocking communication with the instrument
FIX or WORKAROUND
- Remove antivirus software to confirm that the antivirus software is causing the issue
- Configure the antivirus software to not interfere with MassLynx operation (responsibility of customer, not Waters). See WKB21150 for antivirus scanning exclusions to put in place
ADDITIONAL INFORMATION
See the Antivirus software section of note #985: What do I need to do to put my MassLynx PC on a network?