Pump inlet parameters not printing in experimental record nor from MassLynx Inlet Editor - WKB5178
Article number: 5178
SYMPTOMS
Pump inlet parameters including the gradient table, not printing:
- in the experimental record
- from File > Print in the Inlet editor
ENVIRONMENT
MassLynx 4.1 and 4.2
CAUSE
Inlet method created from File > New in the working project
FIX or WORKAROUND
Create all new methods using the default inlet method file:
- Open the Default project.
- Open the default inlet method.
- Save it with a new name (alpha-numeric characters and the underscore only) in the intended project.
- Configure the inlet parameters in the method as needed.