MassLynx communications to mass spectrometer drop off - WKB49052
SYMPTOMS
- No communications to the mass spectrometer
- Intermittent communication problems with mass spectrometer
ENVIRONMENT
- MassLynx
- All instruments
CAUSE
A USB device for internet access has been installed. The device may also have installed its own software, which acts as a DHCP server and conflicts with the Waters DHCP server upon which instrument communications depend. Even after the USB Wi-Fi/3G/4G dongle is removed, if the USB device software remains in place, the conflict with the Waters DHCP server will still cause problems.
Examples of devices and software that are known to cause this problem:
A1 (Austria)
Alcatel Wireless HS-USB WWAN Adapter
ATT USB devices
Linksys AE6000 Wireless-AC Mini USB Adapter
Sierra Wireless Aircard
UMTS stick is HUAWEI Mobile Broadband E173
Unlimited Huawei USB Modem E173
Vodafone USB devices
Telenor software
Vodafone USB Dongle Software
T-Mobile Web'n'walk Manager (ameisvc.exe)
FIX or WORKAROUND
- Shut down MassLynx and run Stop Acquity Processes.
- Remove any USB wireless, 3G, or 4G dongle from the PC.
- Use Windows Control Panel > Programs and Features to remove any software associated with USB internet access devices.
- Reboot the host PC, the EPC, and any ACQUITY components in the system.
- Restart MassLynx.