Error message "Failure to run Analytical Component Engine" in MassLynx - WKB3726
Article number: 3726
SYMPTOMS
- Error message "Failure to run Analytical Component Engine" after system reboot
- Error message "Unable to run external application" might also occur in combination with the one above
ENVIRONMENT
- MassLynx
CAUSE
The error message relates to the Inlet Kernel (the part of the software that controls/defines the LC system). This error can appear if the inlet configuration files or registry keys become corrupted.
FIX or WORKAROUND
- Run MassLynx repair. If error persists, proceed with the following steps:
- Back up MassLynx.
- Uninstall MassLynx.
- Uninstall ICOPs.
- Delete the ICOP folder located at C/Drive > Program Files (x86) > Micromass Utilities.
- Install the current MassLynx SCN and respective ICOPs (WKB6207).
- Reboot the full system, including the PC.
- Establish communication with all instruments before opening MassLynx.
- If error persists, reimage PC.
- Ensure there are no active antivirus software running.
- Repeat steps 5 to 7.
ADDITIONAL INFORMATION
