Cannot connect to Agilent 7890 GC after upgrading to Windows 10 with MassLynx - WKB243169

Article number: 243169

SYMPTOMS

  • Upgraded to Windows 10 and newer version of MassLynx (SCN 1035, SCN 1040)
  • Now cannot control the GC, cannot load a method; it is dimmed/unavailable, the temperature is set to 0 °C and cannot be changed

ENVIRONMENT

  • MassLynx 4.2
  • Windows 10
  • Agilent 7890 GC
  • Xevo TQ-GC
  • Xevo TQ-S micro

CAUSE

The latest versions of MassLynx do not support the 7890 GC after the introduction of support for the Agilent 8890 GC.

FIX or WORKAROUND

  1. Install the last version of MassLynx to support the 7890 as listed in the release notes.
  2. For the Xevo TQ-S micro, this was MassLynx 4.2 SCN 1001, released in July 2019.
  3. For the Xevo TQ-GC micro, this was MassLynx 4.2 SCN 978, released in September 2018.

Note: See article WKB554 for the versions of software to support your MS.

ADDITIONAL INFORMATION

 

id243169, 7890 GC, 7890 GCDET, 7890BGC, A-SQ, A-SQ, MLYNX, MLYNXV41, Q3100MSD, QDA, RADIAN, SCN1001, SCN1035, SCN1040, SCN978, SQD, SQD2, SUPMM, TQD, TQDIVD, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD