TargetLynx does not calculate concentrations - WKB242836
Article number: 242836
SYMPTOMS
- The Concentration Column of the TargetLynx Report is empty
ENVIRONMENT
- MassLynx
CAUSE
Manual integration or peak identification was performed, and the data was not reprocessed within the TargetLynx Browser.
FIX or WORKAROUND
- Save the Modified Peak.
- Click the Process icon.
- Select the box for Quantify. (On newer TargetLynx, you can select "Save Manual integrations".)
- Click OK.
- Verify that the concentrations are now calculated on the report.
ADDITIONAL INFORMATION
id242836, MLYNX, MLYNXV41, SUPMM, TARLYNX41