No calibration or concentration results show in the TargetLynx browser after installing MassLynx on a new workstation - WKB242801
Article number: 242801
SYMPTOMS
- MassLynx installed on new workstation
- Processed first data set with TargetLynx
- No calibration curve or concentration results appear
- Found: Sample Type column is missing from sample list
- Found: Sample Type column added to sample list, Sample Type for standard samples not specified as STANDARD
ENVIRONMENT
- MassLynx
- TargetLynx
- QuanLynx
CAUSE
Sample Type for standard samples not specified as STANDARD
FIX or WORKAROUND
- Set the Sample Type for the standard samples to STANDARD.
- Save the sample list.
- Process the samples with TargetLynx.
ADDITIONAL INFORMATION
