Error: Service Due message - WKB20308
Article number: 20308
SYMPTOMS
- Console displays TQD service due
- No other errors in console logs
- System ran fine previously
ENVIRONMENT
- ACQUITY TQD
- SQD2
- Xevo TQ-S micro
- Xevo TQ-S
- Xevo G2-S
- Xevo TQ-XS
- SYNAPT G2-Si
- MassLynx
- RADIAN ASAP
- Xevo TQ Absolute
- ACQUITY QDa Mass Detector
- SYNAPT XS
CAUSE
The date entered for the last PM was over one year ago.
FIX or WORKAROUND
Disable the service due entry. In the console, do the following:
- Select the MS instrument in the left pane of the MS console.
- Select Configure > Intellistart Configuration.
- Clear the Service due entry check box.
- Click OK to clear the error message.
ADDITIONAL INFORMATION
Note: For the QDa Mass Detector, the option from the MS Console is Configure > Health Monitor Configuration.