OALogin overwrites the SampleID and Data File Name with the Username in SCN 987 and SCN 1036 - WKB203250
Article number: 203250
SYMPTOMS
- Sample ID specified on OALogin page is replaced by whatever is in the Job ID field on the login page—this is usually the Username plus count number and sample number incremented (e.g., AnalystA1-1 and AnalystA1-2)
- In the experimental record and the OALogin report, the Original Sample ID is not recorded
- The MSDataName or MassLynx data file (.raw) is also renamed with the same Username/Job ID reference (and sample number incremented)
- Faults in SCN 987, SCN 1014, and SCN 1036
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Autopurification
- FractionLynx
- MassLynx 4.2
- OALogin
CAUSE
Bug in the software
Faults raised: CRI-3561 and CRI-3562
FIX or WORKAROUND
No longer an issue in SCN 1046 and subsequent releases
ADDITIONAL INFORMATION