Cannot open the chromatogram from the MassLynx sample list - WKB203471
Article number: 203471
SYMPTOMS
- Cannot open a chromatogram by highlighting the row of the sample list and clicking "Chromatogram" at the top
- The chromatogram can be opened from File/Open Data File
- The data is inside the currently open project
ENVIRONMENT
- MassLynx 4.2
CAUSE
There is a space in the project name.
FIX or WORKAROUND
Ensure that there are no spaces or illegal characters in the file names or project names.
ADDITIONAL INFORMATION
See this article for a list of illegal characters: