Error "Failed to obtain detector status information" appears when starting MassLynx analysis with an ACQUITY Column Manager - WKB18280

Article number: 18280

SYMPTOMS

  • Inlet Method window error message:
    Failed to obtain detector status information waiting for method conditions
    Failed to download method :- <<method name>>
  • Losses observed when scanning using NetEvaluator  (See article #1018, What is the NetEvaluator utility?)
  • Column Manager errors in ACQUITY Log
  • Network cables are not faulty

ENVIRONMENT

  • MassLynx
  • ACQUITY Column Manager
  • LC-MS system

CAUSE

Faulty network card in the column manager

FIX or WORKAROUND

Replace the ACQUITY Column Manager Ethernet switch.

ADDITIONAL INFORMATION

MassLynx classifies the Column Manager as a detector.