How to check the accurate mass values used for calibrating MS systems in MassLynx - WKB15244
Article number: 15244
OBJECTIVE or GOAL
Check the accurate mass values used by MassLynx to perform instrument calibration.
ENVIRONMENT
- LCT
- LCT Premier
- LCT Premier XE
- Q-Tof Ultima
- Q-Tof API US
- Q-Tof Premier
- SQ Detector
- SQ Detector 2
- TQ Detector
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- MassLynx
PROCEDURE
- Use Windows Explorer to navigate to the MassLynx folder on the C Drive.
- Open the Ref folder.
- Open the specific reference file for the calibration solution you are using (for example, NaFormate.ref for sodium formate).
- The file contains a list of all the reference peaks.
ADDITIONAL INFORMATION
You can use Notepad.exe to read the .ref files.