"MassLynx Status file not updated" error when starting a run in Chronos with HDX-2 - WKB15510
Article number: 15510
SYMPTOMS
- Error message appears when starting a run in Chronos:
- "An error occurred during the schedule execution. MassLynx Status file not updated in 4327 s"
ENVIRONMENT
- HDX-2
- MassLynx
- Chronos
CAUSE
MassLynx was reinstalled.
FIX or WORKAROUND
- From MassLynx, select Tools > Options.
- In the Options dialog box, change the MassLynx "Refresh rate" from the default of 60 seconds to 2 seconds.
ADDITIONAL INFORMATION