The MassLynx tune page does not recognize the UniSpray - WKB124420
Article number: 124420
SYMPTOMS
- Cannot use UniSpray source with MassLynx
- When I replace the ESI source with a UniSpray source, the tune page does not update to allow selection of the UniSpray
- UniSpray does not show in the Tune Page > Ion Mode drop-down menu
- When swapping from ESI to UniSpray there is no message in MassLynx about the UniSpray source, it is not recognised by the software.
- e.g. MassLynx 4.1 and SCN905 for Xevo TQ and Xevo TQ-S
ENVIRONMENT
- UniSpray
- MassLynx
- Xevo TQ-S
- Xevo TQ MS
- ACQUITY QDa Mass Detector
- SYNAPT XS
- SYNAPT G2-Si
- Xevo TQ-GC
CAUSE
Either this software version or the MS instrument does not support the UniSpray source.
FIX or WORKAROUND
Check to see if your instrument is supported with the UniSpray Source via this article: Is UniSpray supported with my MS instrument? - wkb99192
If it is supported then check the release notes of the latest software version for UniSpray support via this article: What is the latest version of MassLynx software that supports my MS instrument? - WKB554
ADDITIONAL INFORMATION
Also see Cannot operate Xevo TQ-S instrument fitted with UniSpray source enclosure - wkb45025