Is UniSpray supported with my MS instrument? - WKB99192

Article number: 99192

ENVIRONMENT

  • UniSpray
  • MassLynx
  • SQ Detector 2 (SQD2)
  • Xevo TQ Absolute
  • Xevo TQ Absolute XR
  • Xevo TQD
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo G2-XS
  • Xevo G3

ANSWER

Only the instruments listed in the Environment section of this article support use of the UniSpray source. 

UniSpray is NOT supported with ACQUITY QDa Mass Detector, SYNAPT G2-Si, Xevo G2 ToF, Xevo G2 QToF, or Xevo TQ-GC. There are no installation or performance specifications for UniSpray with these instruments.

UniSpray is not officially supported for the SYNAPT XS. It works, but the UniSpray source is not automatically recognized. You must use the "capillary" voltage slider to control the impactor voltage. In reports, the impactor voltage remains labeled as ESI voltage, and all spectra will be labeled ESI instead of UniSpray.

ADDITIONAL INFORMATION

See the release notes for your instrument to determine which version has UniSpray support.
See the article What is the latest version of MassLynx software that supports my MS instrument? - WKB554

Not able to find a solution? Click here to request help.